Model Number 1054096 |
Device Problems
Application Program Problem (2880); No Flow (2991); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/17/2021 |
Event Type
malfunction
|
Event Description
|
The manufacturer received information alleging a ventilator stopped providing flow.There was no harm or injury reported.The device has yet to be returned to the manufacturer for evaluation.At this time, we are unable to confirm the alleged malfunction.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
|
Manufacturer Narrative
|
The manufacturer previously reported receiving information alleging a ventilator stopped providing flow.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed, ventilator inopertaive codes were observed in the ventilator's downloaded event log.The device's software was reloaded to address the issue.The ventilator was returned to the product investigation laboratory for further investigation.During the investigation, the root cause for the malfunction was due to too many files being stored on the sd card.
|
|