Model Number 1054096
Device Problems Application Program Problem (2880); No Flow (2991); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2021
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator stopped providing flow.There was no harm or injury reported.The device has yet to be returned to the manufacturer for evaluation.At this time, we are unable to confirm the alleged malfunction.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer previously reported receiving information alleging a ventilator stopped providing flow.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed, ventilator inopertaive codes were observed in the ventilator's downloaded event log.The device's software was reloaded to address the issue.The ventilator was returned to the product investigation laboratory for further investigation.During the investigation, the root cause for the malfunction was due to too many files being stored on the sd card.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge lane
murrysville PA 15668
MDR Report Key11495255
MDR Text Key240060224
Report Number2518422-2021-00758
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959026469
UDI-Public00606959026469
Combination Product (y/n)N
PMA/PMN Number
K083526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1054096
Device Catalogue Number1054096
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2021
Initial Date Manufacturer Received 02/17/2021
Initial Date FDA Received03/16/2021
Supplement Dates Manufacturer Received04/15/2021
Supplement Dates FDA Received04/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-