MPRI ATTAIN ABILITY STRAIGHT MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 439678
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a device upgrade procedure the existing right atrial (ra) and left ventricular (lv) leads had to also be explanted and replaced due to adhesion between the leads.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the full lead was returned in segments, analyzed.Analysis indicated the outer insulation was ruptured.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The right ventricular (rv) lead that was electively explanted and replaced was returned and analyzed.The rv lead tested out of specification for the outer insulation being ruptured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN ABILITY STRAIGHT MRI SURESCAN
Common Device Name
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key11495919
Report Number2649622-2021-05099
Device Sequence Number19826354
Product Code OJX
UDI-Device Identifier00613994719096
UDI-Public00613994719096
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P080006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/07/2012
Device Model Number439678
Device Catalogue Number439678
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2021
Initial Date Received by Manufacturer 02/26/2021
Supplement Date Received by Manufacturer03/30/2021
Initial Report FDA Received Date03/16/2021
Supplement Report FDA Received Date04/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
W1TR03 CRTP, 5076-52 LEAD, 5076-45 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age47 YR
-
-