| Model Number |
439678 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during a device upgrade procedure the existing right atrial (ra) and left ventricular (lv) leads had to also be explanted and replaced due to adhesion between the leads.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned in segments, analyzed.Analysis indicated the outer insulation was ruptured.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The right ventricular (rv) lead that was electively explanted and replaced was returned and analyzed.The rv lead tested out of specification for the outer insulation being ruptured.
|
| |
|