• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 9/10 COCR HEAD 28MM -3; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 9/10 COCR HEAD 28MM -3; SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1521-90-057
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 02/27/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2005, the primary surgery was performed via tha with the stem, head and liner in question.On (b)(6) 2021, the patient was urgently transported to the hospital and it was found that the head and stem disassociated secondary to wear on the stem.In the revision, the surgeon removed the head, washed the stem, replaced the liner with the new one and implanted the new head.The surgeon commented that the dislocation improved, and the range of motion stabilized.The surgery was completed successfully without any surgical delay.No further information is available.Doi: (b)(6) 2005.Dor: (b)(6) 2021; unknown hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
9/10 COCR HEAD 28MM -3
Common Device Name
SMALL TAPER (9/10) FEMORAL HEADS : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key11496713
Report Number1818910-2021-05094
Device Sequence Number2965406
Product Code JDI
UDI-Device Identifier10603295276623
UDI-Public10603295276623
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K871867
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1521-90-057
Device Catalogue Number152190057
Device Lot Number1882697
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/27/2021
Supplement Date Received by Manufacturer05/04/2021
Initial Report FDA Received Date03/16/2021
Supplement Report FDA Received Date05/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
9/10 COCR HEAD 28MM -3.; ENDURON 10D 50OD X 28ID.; REPLICA A STEM 10.0MM.; UNK HIP LOCKING RING.; 9/10 COCR HEAD 28MM -3; ENDURON 10D 50OD X 28ID; REPLICA A STEM 10.0MM; UNK HIP LOCKING RING
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
-
-