EXACTECH, INC. EQUINOXE; MDLR PRM RMR XLG PLT TIP Back to Search Results
Catalog Number 315-25-15
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pending evaluation.
 
Event or Problem Description
As reported, the surgeon was reaming the glenoid as preparation for the implant.When the surgeon was using his final reamer for the xl glenoid implant, the reamer caught part of a retractor, and broke.This was understood by the surgeon that the reamer was being pushed with significant force, and when it caught a solid piece of the metal the reamer the force being put on the reamer caused it to break.The patient was not injured due to this, and there was only a minimal delay to the surgery.The patient was not harmed in the processed.There were no parts or pieces that fell into the open wound.Patient was last known to be in stable condition following the event.The device will return for evaluation.
 
Additional Manufacturer Narrative
Section h10: (h3) the broken reamer reported in (b)(4) was likely the result of the reamer coming in contact with a retractor during use which led to brittle fracture of the blades.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
MDLR PRM RMR XLG PLT TIP
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
MDR Report Key11496839
Report Number1038671-2021-00113
Device Sequence Number2856612
Product Code KWT
UDI-Device Identifier10885862085375
UDI-Public10885862085375
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
UNK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number315-25-15
Device Lot Number119750002
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/24/2021
Supplement Date Received by Manufacturer03/17/2021
Initial Report FDA Received Date03/16/2021
Supplement Report FDA Received Date04/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-