U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 8; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 8; DXE Back to Search Results
Model Number CAT8XTORQ115-A
Medical Device Problem Codes Material Deformation (2976); Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the deep femoral vein using an indigo system aspiration catheter 8 (cat8) and a non-penumbra sheath.During the procedure, the cat8 did not aspirate and subsequently, the physician noticed the cat8 was flattened in two locations mid-shaft.It was unknown whether the damage occurred upon removal from the packaging or while attempting to aspirate.Therefore, the cat8 was no longer used.The procedure was completed using another cat8 and the same sheath.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned cat8 confirmed that the catheter was ovalized in the mid-shaft.If the device is forcefully gripped, pinched or otherwise mishandled during use, damage such as an ovalization may occur.Further evaluation revealed additional ovalizations throughout the catheter shaft.This damage was incidental to the reported complaint and may have occurred during packaging for return.During the functional test with a demonstration canister and pump max, the cat8 was able to aspirate water into the canister.Penumbra devices are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text: placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 8
Common Device Name
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11497930
Report Number3005168196-2021-00511
Device Sequence Number5257523
Product Code DXE
UDI-Device Identifier00815948022478
UDI-Public00815948022478
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160533
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/09/2023
Device Model NumberCAT8XTORQ115-A
Device Catalogue NumberCAT8XTORQ115
Device Lot NumberH10743
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/17/2021
Supplement Date Received by Manufacturer06/22/2021
Initial Report FDA Received Date03/16/2021
Supplement Report FDA Received Date06/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-