Model Number 186717645
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Failure of Implant (1924); Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional device product codes: osh, mnh, kwp, mni, kwq.Part of the screw remained in the patient¿s bone; device not considered explanted.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2019 the patient underwent a pps to treat the traumatic injury.On (b)(6) 2021 x-rays were taken which revealed that two (2) screws had broken off in the centrum.On (b)(6) 2021 the patient underwent a revision procedure.The fragments were left in the patient's body.Other implants were successfully explanted.No further information is available.This report is for one (1) viper mono screw 6x45mm ti.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: h3, h6: a review of the receiving inspection (ri) for viper mono screw 6x45mm ti was conducted identifying that lot number rl261427 was released in a single batch.Batch1: lot qty of (b)(4) units were released on october 4, 2017 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER MONO SCREW 6X45MM TI
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key11498851
Report Number1526439-2021-00463
Device Sequence Number19957758
Product Code NKB
UDI-Device Identifier10705034123479
UDI-Public(01)10705034123479
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number186717645
Device Catalogue Number186717645
Device Lot NumberRL261427
Initial Date Received by Manufacturer 02/17/2021
Supplement Date Received by Manufacturer04/16/2021
Initial Report FDA Received Date03/16/2021
Supplement Report FDA Received Date05/12/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN LOCKING/SET SCREWS; UNKNOWN RODS; VIPER MONO SCREW 6X45MM TI
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-