| Model Number |
0682400550 |
| Medical Device Problem Code |
Electromagnetic Interference (1194)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was reported that an equipment boom electrical rotational brakes disengaged when a cauterizer device was being used during a procedure.There was patient involvement due to the issue occurring during a procedure but there was no impact to the patient and caused no surgical delay.There were no reported injuries or adverse consequences.All of the equipment booms for this account were manufactured before 22apr2020 and have the older style mfr board.The sfst investigation results can be referenced on sales force complaint (b)(4), the case was closed because we are waiting to performing the pfa inspection.S-series capacitive touch emi disturbances has been investigated on nc/capa (b)(4).Although we couldn¿t confirm the root cause we suspect the issue is most likely due to the previous version of mfr capacitive board.The newer revision capacitive touch board was redesigned and released back in april 2020 to be less susceptible to emi from high frequency devices.A stryker representative will be performing the pfa inspection and will install the current revision of mfr board and confirm that the boom is operational.Once additional information is obtained around this event, a supplemental will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported the brakes are clicking on the s-boom while cautery is in-use in or 11.There were no reported injuries or adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
Field d4 has been updated to include serial number.This boom is confirmed to be within scope of field action pfa2497567 (ref (b)(4)).A stryker representative will be performing the pfa inspection and will install the current revision of mfr board and confirm that the boom is operational.Field g6 has been updated to "follow-up report 001." field h9 has been updated to "additional information" and "correction.".
|
| |
|
Event or Problem Description
|
|
It was reported the brakes are clicking on the s-boom while cautery is in-use in or 11.There were no reported injuries or adverse consequences.
|
| |
|
Search Alerts/Recalls
|