Model Number 657414
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon completion of investigation, additional information will be provided in a supplemental report.
 
Event or Problem Description
As reported: "during a variax 2 plate procedure with a 3.5mm cortical screw (inserted using a hand screwdriver) a metal filament outside the screw or plate was developed.".
 
Event or Problem Description
As reported: "during a variax 2 plate procedure with a 3.5mm cortical screw (inserted using a hand screwdriver) a metal filament outside the screw or plate was developed.".
 
Additional Manufacturer Narrative
Correction: please refer to section d9/h3, the device was not returned.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A potential non-conformity report was nonetheless initiated to address this event.The comprehensive root cause analysis of the potential ncr included a surface and microstructure analysis as well as nano-indent hardness measurement of variax1 and variax2 screw samples by the fraunhofer institute on behalf of stryker.Further internal investigation efforts were focusing on the anodization color difference between variax1 and variax2 screws.Significant differences between the screws which could result in the reported complaints have not been revealed in either fraunhofer or stryker investigations.Excessive in-house handling and bench testing with variax2 screws from various manufacturing batches showed that locking is achieved as intended.The desired increase in torque indicating to the customer that the screw is locking in the plate, can be confirmed.Since the lot number of the device was not communicated, a review of the device history was not possible.A review of the labeling did not indicate any abnormalities.Considering the information given and based on the above investigations a root cause of the reported event could not be determined.If any further substantial information is provided, the investigation report will be updated.H3 other text : device disposition is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L14MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key11499247
Report Number0008031020-2021-00091
Device Sequence Number2776901
Product Code HWC
UDI-Device Identifier07613327073782
UDI-Public07613327073782
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K180500
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number657414
Device Catalogue Number657414
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/19/2021
Supplement Date Received by Manufacturer04/15/2021
Initial Report FDA Received Date03/16/2021
Supplement Report FDA Received Date05/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-