| Model Number |
657414 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "during a variax 2 plate procedure with a 3.5mm cortical screw (inserted using a hand screwdriver) a metal filament outside the screw or plate was developed.".
|
| |
|
Event or Problem Description
|
|
As reported: "during a variax 2 plate procedure with a 3.5mm cortical screw (inserted using a hand screwdriver) a metal filament outside the screw or plate was developed.".
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: please refer to section d9/h3, the device was not returned.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A potential non-conformity report was nonetheless initiated to address this event.The comprehensive root cause analysis of the potential ncr included a surface and microstructure analysis as well as nano-indent hardness measurement of variax1 and variax2 screw samples by the fraunhofer institute on behalf of stryker.Further internal investigation efforts were focusing on the anodization color difference between variax1 and variax2 screws.Significant differences between the screws which could result in the reported complaints have not been revealed in either fraunhofer or stryker investigations.Excessive in-house handling and bench testing with variax2 screws from various manufacturing batches showed that locking is achieved as intended.The desired increase in torque indicating to the customer that the screw is locking in the plate, can be confirmed.Since the lot number of the device was not communicated, a review of the device history was not possible.A review of the labeling did not indicate any abnormalities.Considering the information given and based on the above investigations a root cause of the reported event could not be determined.If any further substantial information is provided, the investigation report will be updated.H3 other text : device disposition is unknown.
|
| |
|
Search Alerts/Recalls
|