ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
MF-410375 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Granuloma (1876)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation:visual analysis of the returned device identified: weight to the spec, crease fold, deformation.Cloudy and bubbles were observed after autoclave disinfection process.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.The event of "granuloma_" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: granuloma.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "pain and discomfort" and implant exchange from textured to smooth breast implants due to the patient¿s concern with the product.Additionally, healthcare professional reported "fibroconnective tissue with foreign body granulomatous reaction consistent with capsule." device has been explanted.
|
| |
|
Search Alerts/Recalls
|
|
|