| Model Number |
TRLP-440 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
02/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iii.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified; fold creases, wear abrasion, weight within specification.Based on the device analysis the final assessment is: no issues found related to the manufacturing process.The crease/folding of implant condition that was indicated in the complaint was observed, nevertheless is not considered a workmanship, since the device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional additionally reported left side "any creases" and "double capsule." device has been explanted.
|
| |
|