MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINN LNR CON +4 10D 28IDX48OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINN LNR CON +4 10D 28IDX48OD; PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS Back to Search Results
Model Number 1218-28-748
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Lip of constrained 10 degree poly liner broke when the hip was being reduced.Reason for the revision was dislocation.Surgery delayed due to the reported 15 minutes.
 
Event or Problem Description
Additional information was received stating that there was no any adverse consequences that affected the patient because of the reported event.This event was an intra-op fracture and happened during revision surgery.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Removed codes joint dislocation, device revision or replacement and implant dislocation (device dislodged or dislocated).The device code fracture is reported again to capture the updated code implant fracture intra-op: polyethylene.
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4) investigation summary
=
> no device associated with this report was received for examination.A search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot
=
> a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Device history review
=
> a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN LNR CON +4 10D 28IDX48OD
Common Device Name
PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11506191
Report Number1818910-2021-05259
Device Sequence Number14147385
Product Code KWZ
UDI-Device Identifier10603295011521
UDI-Public10603295011521
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K043058
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1218-28-748
Device Catalogue Number121828748
Device Lot NumberHR0407
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/01/2021
Supplement Date Received by Manufacturer03/17/2021
04/08/2021
Initial Report FDA Received Date03/17/2021
Supplement Report FDA Received Date04/01/2021
04/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINN LNR CON +4 10D 28IDX48OD; UNKNOWN HIP FEMORAL HEAD; PINN LNR CON +4 10D 28IDX48OD; UNKNOWN HIP FEMORAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
-
-