NOVA BIOMEDICAL CORP NOVA STATSTRIP GLUCOSE HOSPITAL METER SYSTEM; BLOOD GLUCOSE METER Back to Search Results
Model Number 53398
Medical Device Problem Code Patient Data Problem (3197)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no reported patient harm or intervention.There is currently a pending investigation.A request has been sent to the customer obtain meters and barcode or sample of a test barcode for verification of 1-d versus 2-d barcodes.Additional information and further details will be provided in a supplemental report.
 
Event or Problem Description
A customer reported that gluc conn meters (1.75) mis-scanning 1d bar showing incorrect patient id's.The account has 1 & 2 d bar codes on patients wrists.There has been no report of patient harm or adverse event.
 
Additional Manufacturer Narrative
An investigation has been initiated and is currently pending.Further details will be provided in an additional supplemental report when they become available.
 
Additional Manufacturer Narrative
Additional and corrected information: b4, d9, g6, h1, h2, h3, h10.An investigation has been initiated and is currently pending.Further details will be provided in an additional supplemental report when they become available.
 
Additional Manufacturer Narrative
Udi: (b)(4).The customer reported an incorrect account number shown when scanned by the statstrip 1.75 meter (b)(6).This occurred on (b)(6) 2021.The meter scanned a patient identification number as and accessed the incorrect patient information.The customer states there was not patient harm as a result of this.The meter was returned to nova biomedical corp.For investigation.The scanner window was free of damage or debris and seemed to not effect the scanner on the meter.An unused barcode was provided by the customer which was scanned fifty times without producing an error.The customer did not return one of the malfunctioning barcodes for investigation.The barcode was put through a lvs-9570 barcode verifier which gave the barcode a grade of "f" for decodability and modulation.A device history record (dhr) review for the meter was performed.The review included an assessment of the production, testing, and release of the products.No abnormalities or concerns were observed; the dhr indicated the released product met all specifications.Previously opened and investigated incidents with similarily-graded barcodes determined the root cause for the patient misreads to be attributed to the user facility using poor quality 1d barcodes that were printed using code -39 symbology.Code-39 is no longer industry standard and has a higher substitution failure rate than code-128 and much higher than the standard 2d barcodes available.This failure mode does not represent a systemic issue or a failure of the meter to perform as intended.The failure mode is directly related to the quality of the barcode being generated by the customer.The root cause of the misread is most likely due to poor barcode quality.Based on the original investigation, it is suggested to prevent barcode mis-reads in the future, the customer should change from a 1d to a 2d barcode, which would greatly reduce the chances of a barcode misread.Trending will be monitored for this and or similar complaints.A corrective and preventative action (capa) is not required as a result of this investigation because the root cause of the reported issue was determined to be poor barcode quality.This is report 1 of 2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVA STATSTRIP GLUCOSE HOSPITAL METER SYSTEM
Common Device Name
BLOOD GLUCOSE METER
Manufacturer (Section D)
NOVA BIOMEDICAL CORP
200 prospect street
waltham MA 02454 9141
MDR Report Key11507527
Report Number1219029-2021-00009
Device Sequence Number4266267
Product Code CGA
UDI-Device Identifier00385480533980
UDI-Public00385480533980
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K063821
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation 505
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number53398
Device Catalogue Number53398
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2021
Supplement Date Received by Manufacturer02/18/2021
02/18/2021
06/10/2021
Initial Report FDA Received Date03/17/2021
Supplement Report FDA Received Date04/14/2021
05/13/2021
06/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-