| Model Number |
25AGFN-756 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Biocompatibility (2886)
|
| Health Effect - Clinical Codes |
Thrombosis/Thrombus (4440); Valvular Insufficiency/ Regurgitation (4449)
|
| Date of Event |
02/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021, a patient had undergone an implant procedure of a 25mm regent.On (b)(6) 2021, the patient was brought back for an emergent redo avr procedure.One of the leaflets of the regent valve was not pivoting properly, this caused a large jet to come through, and one leafet appeared to not open.The leaflet appeared to have thrombosis on it.Dr.(b)(6) replaced this valve with an edwards magna ease.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of the one of the leaflets of the regent valve not pivoting properly was reported.Morphological and hydrodynamic examination indicated the valve met abbott specifications.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Hydrodynamic testing upon return to abbott indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The cause of the reported event could not be conclusively determined.
|
| |
|