• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH AO/ASIFA REAMING ATTACHMENT F/BPL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH AO/ASIFA REAMING ATTACHMENT F/BPL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.780
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The reporter¿s phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
 
Event or Problem Description
It was reported from (b)(6) that during sterilization, it was observed that the reaming attachment device became dismantled.The event was not related to surgery.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the device did not function, had a component damage, and the moving parts did not move smoothly.It was further determined that the device failed pretest for check the tool coupling and check the free movement.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AO/ASIFA REAMING ATTACHMENT F/BPL
Common Device Name
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key11509502
Report Number8030965-2021-01897
Device Sequence Number20170812
Product Code GEY
UDI-Device Identifier07611819137219
UDI-Public07611819137219
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number530.780
Device Lot Number3040925
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2021
Supplement Date Received by Manufacturer05/07/2021
Initial Report FDA Received Date03/17/2021
Supplement Report FDA Received Date05/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-