| Model Number |
MEDIUM-HD-G1 |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure from normal wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported by (b)(6) that during service and evaluation, it was determined that the attachment device exhibited heat.It was observed that the device had bearing damage and the labeling had an illegible etch.It was further determined that the device failed pretest for labeling assessment and temperature verification.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
| |
|