• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. MULTIVAC 50 XL ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. MULTIVAC 50 XL ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ASC4730-01
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/12/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during a hip procedure, the wand pieces fell off.The piece was removed from the patient using grasping forceps.Backup device was available to complete the procedure.No significant delay and no patient complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection of the returned instrument shows portion of the electrode is detached.The device was plugged into the controller and registered settings (1,7).The wand was able to generate plasma as intended.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.The complaint was confirmed.Factors that could have contributed to the reported event include: (1) hitting the device tip against a hard surface.(2) using the device as a lever to enlarge surgical site.(3) mechanical displacement of tissue through applied force.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIVAC 50 XL ICW
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key11509983
Report Number3006524618-2021-00332
Device Sequence Number19794475
Product Code GEI
UDI-Device Identifier00817470006261
UDI-Public00817470006261
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K033584
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/13/2022
Device Model NumberASC4730-01
Device Catalogue NumberASC4730-01
Device Lot Number2031910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2021
Initial Date Received by Manufacturer 02/25/2021
Supplement Date Received by Manufacturer04/14/2021
Initial Report FDA Received Date03/17/2021
Supplement Report FDA Received Date04/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age28 YR
Patient Weight68
-
-