SMITH & NEPHEW, INC. LAG SCREW DRIVER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71674067
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/24/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during surgery of meta tan nail removal, the mating part on the lag screw broke.Then, they tried to continue using lag screw driver, inner rod bent and outer rod deformed.Then the surgeon drilled bone around the lag screw drilled and tried to removing a lag screw with a pench.The lag screw could not removed.They gave up removing the lag screw and the nail.3 hours surgical delay.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAG SCREW DRIVER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11510086
MDR Text Key240556559
Report Number1020279-2021-02211
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560605
UDI-Public03596010560605
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674067
Device Catalogue Number71674067
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received06/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-