STRYKER ORTHOPAEDICS-MAHWAH TRITANIUM BPLATE TRIATHLON S5; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Model Number |
5536-B-500 |
| Medical Device Problem Codes |
Loss of Osseointegration (2408); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/22/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An event regarding loosening and subsidence involving a triathlon baseplate was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right knee was revised due to loosening and subsidence of the tibial baseplate.A size 5 tritanium tibial baseplate and unknown liner were revised to a universal baseplate with stem and new liner.There are no allegations against the revised liner.
|
| |
|
Search Alerts/Recalls
|
|
|