• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. AMSCO 400 STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. AMSCO 400 STERILIZER Back to Search Results
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2021
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite to inspect the unit and found the viewport had been damaged allowing water and steam to leak from the sterilizer.The technician replaced the view port, tested the unit, confirmed it to be operating according to specifications, and returned it to service.No additional issues have been reported.
 
Event Description
The user facility reported water and steam leaking from their amsco 400 sterilizer onto the floor.No report of injury.
 
Manufacturer Narrative
A steris service technician collected water samples to be tested.The facility's incoming water was tested, and the results identified that four of the critical chemical attributes of water quality were above the maximum requirements for the electric steam generator as stated in the amsco 400 sterilizer operator manual (table 5-1.Required feed water quality for carbon-steel steam generators).The user facility is responsible for ensuring that their incoming water supply remains within the electric steam generator's operating requirements.The amsco 400 sterilizer operator manual states (3-1), "water quality - supplied must be within specifications.Improper water quality adversely affects equipment operation." the technician notified the customer of the water quality test results and that their water quality is not meeting the required operating conditions.The user facility has committed to making the necessary improvements to the facility's incoming water quality.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMSCO 400 STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
MDR Report Key11514646
MDR Text Key244845930
Report Number3005899764-2021-00011
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/25/2021
Initial Date FDA Received03/18/2021
Supplement Dates Manufacturer Received02/25/2021
Supplement Dates FDA Received04/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-