SYNTHES GMBH T-PAL ADVANCED APPLICATOR HANDLE; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
| Catalog Number |
03.812.520 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from (b)(6) reports an event as follows: it was reported that on (b)(6) 2021 during surgery there were issues implanting the tpal-cages because the applicator was blocked.The cage was placed with some difficulty and the surgery was completed successfully with a ten (10) minute surgical delay.Concomitant devices: unknown applicator knob (part# unknown, lot# unknown, quantity 1).Unknown cage (part# unknown, lot# unknown, quantity 1).This report is for one (1) t-pal advanced applicator handle.This is report 1 of 2 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6: the complaint condition could not be confirmed for the returned device.There was no indication that a design or manufacturing issue has caused the issue and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: part: 03.812.520, lot: 4l98422, manufacturing site: haegendorf, release to warehouse date: 04.July 2019.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|