Catalog Number ROB00036
Device Problem Application Program Problem (2880)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/24/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference number: case-(b)(4).
 
Event Description
It was reported that, during a navio demo, a message saying: "an error occurred during initialization: system memory error (errcode=8)" pop up after rebooting.All the cables from cpu were reset and the machine was rebooted; but the error persisted.As incident occurred during demo, there was no patient involvement.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the navio cpu, pn 220001, sn (b)(6), intended for use in treatment, was returned for evaluation.The reported event was not confirm visually or functionally.The device was connected to a navio console, and advanced to the administrator screen for testing.The device passed testing and operated as design.No error message or malfunction was observed.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, factors that may have contributed to the reported symptom may have been associated with failure of the motherboard and/or handpiece.A historical capa, nc, hhe/pra, field action review was completed.A review of prior escalation actions found no actions applicable to the scope of this case.The failure mode and associated risk have been anticipated within the risk file and that the documented risk level is still adequate.No further containment or corrective action is recommended or required at this time, all complaints are monitored through complaint investigation and post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO SURGICAL SYSTEM INDIA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11515607
MDR Text Key240654771
Report Number3010266064-2021-00206
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberROB00036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/24/2021
Initial Date FDA Received03/18/2021
Supplement Dates Manufacturer Received02/08/2022
Supplement Dates FDA Received02/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKN NAVIO ROBOTICS INSTR/ SERIAL: (B)(6).
-
-