NIHON KOHDEN CORPORATION GZ-120PA; TRANSMITTER Back to Search Results
Model Number GZ-120PA
Medical Device Problem Codes Power Problem (3010); Intermittent Loss of Power (4016)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The biomedical engineer reported that the gz telemeter's battery contact was broken, causing the telemeter to lose power intermittently.
 
Event or Problem Description
The biomedical engineer reported that the gz telemeter's battery contact was broken, causing the telemeter to lose power intermittently.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GZ-120PA
Common Device Name
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 116-8 560
JA  116-8560
MDR Report Key11516611
Report Number2080783-2021-00157
Device Sequence Number3625131
Product Code DRT
UDI-Device Identifier04931921117392
UDI-Public04931921117392
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 03/18/2021
Report Date (Section F) 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGZ-120PA
Device Catalogue NumberGZ-120PA
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date02/24/2021
Device Age38 MO
Event Location Hospital
Date Report to Manufacturer03/18/2021
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA03/18/2021
-
-