Model Number 1049037
Medical Device Problem Codes Disconnection (1171); Mechanical Problem (1384); Unintended Movement (3026)
Health Effect - Clinical Code Low Oxygen Saturation (2477)
Date of Event 01/19/2021
Type of Reportable Event Serious Injury
Event or Problem Description
The manufacturer received information alleging that while on a bilevel positive airway pressure (bipap) device, a patient became confused and removed the performax mask.The patient subsequently required intubation due to oxygen desaturation.The manufacturer has made numerous attempts to contact the reporting facility to gather additional information and to request return of the mask for investigation.No further information has been provided, and no product has been returned.The performatrak full face mask is intended to provide an interface for application of cpap or bi-level therapy to patients.The mask is for single use in the hospital/institutional environment.The mask is to be used on patients (>66lbs/30kg) for whom cpap or bi-level therapy has been prescribed.Labeling warns the user "this mask is not suitable for providing life support ventilation", and that "this mask should not be used on patients who are uncooperative, obtunded, unresponsive, or unable to remove the mask." as no further information was provided, and no product returned, the manufacturer is unable to determine the root cause of the reported event.The manufacturer concludes no further action is necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORMATRAK MASK
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS INC
1001 murry ridge lane
murrysville PA 15668
MDR Report Key11517382
Report Number2518422-2021-00679
Device Sequence Number14436080
Product Code BZD
UDI-Device Identifier00606959000759
UDI-Public00606959000759
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K091271
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Model Number1049037
Device Catalogue Number1049037
Device Lot Number46782941
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2021
Initial Report FDA Received Date03/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-