| Model Number |
PL548S |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/17/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Manufacturing site evaluation: investigation on-going.Should relevant additional information investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap (b)(4) that a atr.Vascular clip 25mm cvd.3.43n arteri.(part # pl548s) was used during an unknown procedure performed on (b)(6) 2021.According to the complainant, the patient suddenly dropped out while grasping the blood vessel with the device.Due to the occurrence of this event, the patient was transferred to open surgery.There was no infection and no health hazard to patient reported.No information was provided as to a specific device malfunction.Although requested, additional information has not been made available.The adverse event is filed under aag reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: visual investigation: vigilance investigator carried out the pictorial documentation visually.Investigation was carried out by q-coordinator of the production plant.No production related trouble found.The error described could be usage related.Batch history review: due to the fact that neither an article number nor a lot number was provided, a review of the device history records must remain incomplete.Conclusion and measures / preventive measures based on the investigation of the production plant, a production related error can be excluded.The closing force test as well as the water bag test was passed.A usage related error cannot be excluded.Based upon the investigations results a capa is not necessary.
|
| |
|
Search Alerts/Recalls
|