• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL CT BRD VIO 27IN 0 S/A CTX; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL CT BRD VIO 27IN 0 S/A CTX; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number J364H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Events reported in (b)(4), 2210968-2021-02500, 2210968-2021-02501, 2210968-2021-02502, 2210968-2021-02503 attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device batch number pdk402, and no non-conformances were identified.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that a patient underwent an ob-gyn procedure on (b)(6) 2021 and suture was used.During the procedure, the suture broke while trying to close the uterine vessel.There were no adverse patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Date sent to the fda: 05/06/2021.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Investigation summary: according to the information received, open hysterectomy whilst tying & closing uterine vessel with 0 vicryl j346h suture material snapped.The product was returned to ethicon inc for evaluation.Visual inspection and functional test evaluation were conducted on the returned device.Visual analysis of the returned sample determined that eleven unopened samples of product code j364h was received for evaluation.Visual inspection of the unopened samples no defects were found on the package.The samples were opened, and the swage and attachment area was noted to be as expected.The sutures were dispensed without problems and examined along the strand and no defects, damage, or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.This report is not intended to deny that you experienced a problem with the device.There may have been other circumstances or issues that occurred during the use of the device that we were unable to duplicate during our laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD VIO 27IN 0 S/A CTX
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08876
MDR Report Key11518702
Report Number2210968-2021-02499
Device Sequence Number14271862
Product Code GAM
UDI-Device Identifier10705031039315
UDI-Public10705031039315
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberJ364H
Device Catalogue NumberJ364H
Device Lot NumberPDK402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2021
Initial Date Received by Manufacturer 03/02/2021
Supplement Date Received by Manufacturer04/12/2021
Initial Report FDA Received Date03/18/2021
Supplement Report FDA Received Date05/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-