| Model Number |
20-260 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/27/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iv.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified the weight of the device within the specification, a deformation, fold crease, wear abrasion, and crystalline appearance.Clouds were observed in the gel after the autoclave disinfection process.Based on the device analysis, the final assessment is no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iv.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|