| Catalog Number |
8065751577 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported that when the trocar and the chandelier got separate, the metal cylinder part came off of the blue plastic part in the trocar during surgery.It was not a problem since the part came off with the chandelier.The product was replaced and procedure completed with no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned trocar cannula/hub assembly was visually inspected and was found to be non-conforming, the hub and cannula were separated.There was presence of adhesive inside of the hub.The product was processed and released according to the product¿s acceptance criteria.During assembly of the trocar product adhesive is dispensed at the cannula hub interface.The evaluation of the returned sample identified adhesive inside of the hub, therefore, how and when the trocar cannula and hub became separated cannot be determined from this evaluation and the root cause for this complaint is unknown.A potential contributing factor of the separation is manipulation during surgery.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.Assemblies are 100% inspected for adhesive application during assembly.If adhesive application is incorrect the assembly is scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|