| Model Number |
72202961S |
| Medical Device Problem Codes |
Crack (1135); Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident. a visual inspection found distal tip and fiber damage, a cracked distal lens, and scratches on the needle. a review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution. a complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee arthroscopy the scope had a blurry/foggy image.The procedure was completed with a backup device and no delay or patient injuries were reported.Results of investigation have concluded that this unit had a cracked distal lens which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|