• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. VIDEOARTHROSCOPE HD 4MM X 30 DEG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. VIDEOARTHROSCOPE HD 4MM X 30 DEG Back to Search Results
Model Number 72202961S
Medical Device Problem Codes Crack (1135); Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).The reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident. a visual inspection found distal tip and fiber damage, a cracked distal lens, and scratches on the needle. a review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution. a complaint history review concluded this was a repeat issue.The complaint was confirmed.Factors that could have contributed to the reported event include an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
 
Event or Problem Description
It was reported that during a knee arthroscopy the scope had a blurry/foggy image.The procedure was completed with a backup device and no delay or patient injuries were reported.Results of investigation have concluded that this unit had a cracked distal lens which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOARTHROSCOPE HD 4MM X 30 DEG
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key11520757
Report Number3003604053-2021-00100
Device Sequence Number4047577
Product Code HRX
UDI-Device Identifier00885554024203
UDI-Public00885554024203
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K043395
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72202961S
Device Catalogue Number72202961S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2021
Initial Report FDA Received Date03/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-