| Model Number |
71631434 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
the device, intended for use in treatment was returned for evaluation: a visual inspection was performed and confirmed the rotation bar has gouges/scratches on the tip.This device also show signs of excessive wear/usage.The manufactured date for this device is 2015.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.& nbsp; & nbsp; .
|
| |
|
Event or Problem Description
|
|
It was reported that, the handle of a meta-tan anti rotation bar is out of alignment.Incident occurred during a meta tan nailing case with instruments outside the patient.The procedure was successfully completed using a back-up device.No other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device, intended for use in treatment was returned for evaluation: a visual inspection was performed and confirmed the rotation bar has gouges/scratches on the tip.This device also show signs of excessive wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the complaint history on the listed part revealed no prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship between the device and the reported incident was corroborated.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|