TERUMO CARDIOVASCULAR SYSTEMS CORPORATION HOLLOW FIBER OXY WITH 4000 ML; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*RX15RW40
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2021
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the temperature probe was not reading temperatures.No patient involvement.The product was changed out.Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 19, 2021.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date).G3 (date received by manufacturer).G6 (indication that this is a follow-up report).H2 (follow-up due to additional information and device evaluation).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 213, 67).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 213 - no device problem found.Investigation conclusions: 67 - no problem detected.The affected samples were inspected upon receipt with no visual anomalies noted.The thermistors on the samples were checked for functionality and they were found to function appropriately.They were also sent for x-ray analysis to confirm that they were fully intact.Representative retention samples from the same product code lot numbers were subjected to multiple sterilization cycles and environmental conditioning cycles and maintained thermistor functionality.All capiox units are 100% visually inspected at several points in the production process.All available information has been placed on file in quality management for appropriate tracking, trending and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOLLOW FIBER OXY WITH 4000 ML
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key11525879
MDR Text Key245522597
Report Number1124841-2021-00059
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450141
UDI-Public(01)00699753450141
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model Number3CX*RX15RW40
Device Catalogue NumberN/A
Device Lot NumberYC17
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-