|
It was reported that during a surgery, the qfix drill and the drill guide were put together correctly.These instruments were inserted into the glenoid but before engaging with the bone, the drill was tested to see if it was rotating by activating the drill but it immediately seized, locked and snapped the drill bit.The product break inside the patient but the pieces were successfully removed.The procedure was successfully completed without a significant delay using a back-up device.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
H10 h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|