• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS SMART PORT; PORT & CATHETER,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS SMART PORT; PORT & CATHETER, Back to Search Results
Model Number CT80STPD-NFVI
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device has yet to be returned to the manufacturer for a device evaluation.An investigation into the root cause for product problem is currently in progress.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event or Problem Description
An end user reported an issue with a smartport ct.During removal of a poorly functioning port, a crack/split in the catheter was discovered.The catheter on the port was split at the 14 cm mark.Port failed and needed replaced.There was no report of patient harm or adverse event by the end user.It was indicated the reported device is available for return to the manufacturer for a device evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation was one port with catheter tubing and blue boot.As received, the catheter and blue boot were attached to the port.The port and catheter tubing contained biomaterial {blood}inside and out.During sample review, the catheter was found to be fractured between the 13 and 14 cm marks.Based on the location of the fracture, about 5.5 cm from the port body, the most likely root cause of the catheter fracture is pinch-off syndrome.This is cautioned in the directions for use (dfu) of the device.The customer's complaint description of crack/split in the catheter is confirmed.The most likely root cause of the catheter fracture is pinch-off syndrome.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: the instructions for use, (16608102-01) which is supplied to the user with this item number, contains the following statements: absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.A blood return should be present prior to usage of device for any therapy or testing.If the patient complains of pain, or if there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Caution: avoid piercing catheter with suture needle.Potential complications: catheter fragmentation and catheter pinch-off.Catheter placement considerations: warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Pinch-off syndrome: pinch-off syndrome signs may include difficulty in aspirating blood, resistance to flushing or infusion of medications or fluids that improves with position changes, infraclavicular pain and/or swelling with catheter flushing or infusion palpitations, sudden onset chest pain, cardiac arrhythmias, extra heart sound, chest wall swelling at the port pocket, vein insertion site, pain in shoulder or port area not associated with swelling, cough, paresthesia of arm on side of catheter withdrawal occlusion or swishing sound with catheter flushing.Warning: avoid medial catheter placement into subclavian vein through percutaneous technique.This placement could lead to catheter occlusion, damage, rupture, shearing, or fragmentation due to compression of the catheter between the first rib and clavicle.Catheter shearing has been reported when the catheter is inserted via a more medial route in the subclavian vein.Note: if infusion or aspiration is successful upon lifting arm above the head and turning the head, consider pinch-off syndrome as a possible cause.The line should be radiologically evaluated if pinch-off syndrome is suspected.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT
Common Device Name
PORT & CATHETER,
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls, ny 12801 NY 12801
MDR Report Key11528410
Report Number1317056-2021-00068
Device Sequence Number11896311
Product Code LJT
UDI-Device IdentifierH787CT80STPDNFVI1
UDI-PublicH787CT80STPDNFVI1
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K101017
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model NumberCT80STPD-NFVI
Device Catalogue NumberCT80STPDNFVI
Device Lot Number5630111
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2021
Supplement Date Received by Manufacturer03/08/2021
Initial Report FDA Received Date03/19/2021
Supplement Report FDA Received Date06/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-