| Model Number |
DYND04131 |
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
02/11/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was reported that the patient was noted to have experienced "abnormal bleeding." the reporting facility indicated that some of their gomco clamps were noted to have missing washers, washers that were "embedded and attached, and washers that were "free floating." the patient required the application of avitene to the procedural site to address the reported abnormal bleeding.No further patient impact, medical treatment, or follow-up care was reported.A sample was returned by the reporting facility and sample investigation is ongoing.Due to the reported need for medical treatment, this medwatch is being file.If additional relevant information becomes available a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was noted to have experienced "abnormal bleeding.".
|
| |
|
Additional Manufacturer Narrative
|
|
Multiple samples were returned to the manufacturer and evaluation of samples was completed on (b)(6) 2021.Inspection of the returned samples found that some gomco clamps were missing washers.During evaluation of the samples, it was noted that clamp components appeared to be mixed from different manufactures.The dynd04131 gomco clamp is a sterile single-use device that is not intended to undergo further processing.It was determined that the gomco clamps are likely being re-processed and re-assembled which is resulting in missing washers and missing clamp components.No lot number for the device involved in the incident has been provided.If additional relevant information becomes available another supplemental medwatch will be filed.
|
| |
|
Search Alerts/Recalls
|