• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG PLUG F/TIBIAL PLATEAU; KNEE ENDOPROSTHESES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG PLUG F/TIBIAL PLATEAU; KNEE ENDOPROSTHESES Back to Search Results
Model Number NN260P
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Joint Laxity (4526)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation results: reference code nx051z, device name as vega ps tibial plateau cemented t1, serial number n/a, batch number 52106248, manufacturing date 09.02.2015.Reference code device name corresponding internal notification number: nx009z as vega ps femoral comp.Cemented f4n l 400479820, nx043 patella 3-pegs p3 400479821, nx110 vega ps gliding surface t1/1+ 10mm 400479822, nn260p plug f/tibial plateau 400479823.Investigation: no products available.Therefore a failure description at the products are not possible.Pictorial documentation there are no pictures avail:able.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all available lot numbers and the products found to be according to our specification valid at the time of production.There are 1 similar complaints against the same lot number(s).Lot number 52082622: (b)(4).Explanation and rationale: no products available and therefore it is hardly possible to determine an exact conclusion and root cause.The exact cause cannot be determined.Corrective action: for this topic (implant loosening) a product safety case (psc) was initiated.
 
Event or Problem Description
It was reported to aesculap inc.That a vega knee implant system was implanted during a primary surgery performed on (b)(6) 2015.According to the complainant, following the primary surgery, the patient experienced loosening of the implant.The patient underwent a revision surgery occurred on (b)(6) 2017.The complaint device was not available to be returned to the manufacturer for evaluation.All available information has been provided at this time, if additional information becomes available the complaint will be updated accordingly.The adverse event / malfunction is filed under reference xc (b)(4).Involved components: nx009z (ps femur cemented f4n lt), nx043 (universal patella p3), nx110 (ps pe insert t1/t1+, 10mm), nn260p (peek plug f/ tibia), nx051z (ps tibia cemented t1).The cement used is unidentified.Associated medwatches: 2916714-2020-00329, 2916714-2020-00330, 2916714-2020-00331, 2916714-2020-00333.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLUG F/TIBIAL PLATEAU
Common Device Name
KNEE ENDOPROSTHESES
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key11528973
Report Number2916714-2020-00332
Device Sequence Number12099771
Product Code JWH
UDI-Device Identifier04046963711320
UDI-Public4046963711320
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/19/2021
Report Date (Section F) 03/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model NumberNN260P
Device Catalogue NumberNN260P
Device Lot Number52093075
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date07/20/2018
Event Location Hospital
Date Report to Manufacturer07/20/2018
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA03/19/2021
-
-