Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Failure of Implant (1924)
Event Type  malfunction  
Manufacturer Narrative
The device was not available for evaluation as it remains in the patient.Imaging provided shows at least one locking cap to have come loose at l4.The rod has translated significantly in the caudal direction, indicating both caps on that side are loose.The exact cause of the reported issue could not be determined.
 
Event Description
It was reported by a representative from italy that the locking cap came out 3 months post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CREO
Type of Device
CREO MIS LOCKING CAP
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon 19403
6109301800
MDR Report Key11532052
MDR Text Key242752957
Report Number3004142400-2021-00050
Device Sequence Number1
Product Code MNH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received02/21/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-