It was reported that during npwt utilizing a renasys foam filler kit, air leakage occurred.It was confirmed that there was a damage on the softport.Applying the fixation strip did stop the leak.But, the soft port was replaced and treatment was continued.There was no patient harm.No delay was reported.
|
H3, h6: the device used in treatment was not returned for evaluation, all provided information has been reviewed and we have not been able to establish a relationship between the reported event or determine a root cause.Probable root cause is a damaged component.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.The complaint history file contains further instances.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.Additional information: d4, h4.
|