MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. SMALL NPWT FOAM FILLER KIT; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. SMALL NPWT FOAM FILLER KIT; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801088
Device Problems Gas/Air Leak (2946); Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/26/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during npwt utilizing a renasys foam filler kit, air leakage occurred.It was confirmed that there was a damage on the softport.Applying the fixation strip did stop the leak.But, the soft port was replaced and treatment was continued.There was no patient harm.No delay was reported.
 
Manufacturer Narrative
H3, h6: the device used in treatment was not returned for evaluation, all provided information has been reviewed and we have not been able to establish a relationship between the reported event or determine a root cause.Probable root cause is a damaged component.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.The complaint history file contains further instances.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.Additional information: d4, h4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL NPWT FOAM FILLER KIT
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11536873
MDR Text Key241357308
Report Number8043484-2021-00659
Device Sequence Number1
Product Code OMP
UDI-Device Identifier04582111152978
UDI-Public4582111152978
Combination Product (y/n)N
PMA/PMN Number
K110647
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Catalogue Number66801088
Device Lot Number2019120299
Was Device Available for Evaluation? No
Date Manufacturer Received05/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-