| Model Number |
CSE-P-225 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Unspecified Blood or Lymphatic problem (4434)
|
| Date of Event |
03/02/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A haemonetics field service engineer evaluated the cell saver 5 autologous blood recovery system.The haemonetics field service engineer did not confirm any defects with the pump platen arm which remained secure with no errors during functional check.The suction pump was replaced during service from haemonetics field service engineer but this was due to the suction pump being noisy.The cell saver® elite set - 225ml sample was discarded by customer and will not be returned to haemonetics for evaluation, without the physical sample haemonetics is unable to perform further evaluation.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2021 haemonetics was notified of a pump platen issue during a procedure in (b)(6), utilizing the cell saver 5 autologous blood recovery system and cell saver® elite set - 225ml.There was a delay in the procedure which required the patient to receive a blood transfusion.
|
| |
|