Manufacturing review of the device history record for the reported lot shows that all units were quality released on 12/10/2020 having met all internal manufacturing specifications and qc acceptance requirements for workmanship.All sterilization processing records and bioburden testing indicated a successful sterilization cycle and passing results of product lal and sterility samples allowing the subassembly lot to be released for further packaging and labeling having met all criteria for release.There were no non-conformance's during manufacturing or sterilization potentially impacting the final acceptance of this manufacturing lot.There were no deviations identified in this lot history record.In lieu of a request for the oem supplier dhr review, it is noted that aziyo processes the non-sterile envelope materials including cutting, suturing, packaging and sterilizing the product.The instructions for use for the cangaroo envelope ((b)(4)) in use at the time of product shipment lists risk of hematoma and infection as potential complications.No additional details are available, should any additional information be obtained regarding this event, a follow-up report will be filed.
|