| Model Number |
86700 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Visual inspection of the device showed damage on the shaft near to the ring # 20.It looks like it was burnt with a external object because no damages were found on the signal wires.Electrical continuity test was performed, no open circuit and no short circuits were found.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.Based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications.
|
| |
|
Event or Problem Description
|
|
Reportable based on analysis completed on 03 mar 2021.It was reported that during preparation of a blazer dx-20 the electrode did not work.The procedure was completed by exchanging the catheter.However analysis revealed catheter shaft damage.
|
| |
|