| Model Number |
0520161A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
| |
|
Event or Problem Description
|
|
A user facility registered nurse (rn) reported that a dialyzer leaked immediately after the initiation of the patient¿s hemodialysis (hd) treatment.The rn visually observed that the dialyzer leaked from two locations.The first leak occurred two inches below the header cap of the dialyzer.The second leak occurred within a fiber.The dialyzer leaks were confirmed to be internal.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.There was no defect or damage seen on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 100 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be available to be returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h3 plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product non-acceptance or deviation in the manufacturing process related to the complaint event.This includes non-conformances, rework, labeling, process controls, and any other occurrence in production that was potentially related to the complaint.The lot met all release criteria.A definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.
|
| |
|
Event or Problem Description
|
|
A user facility registered nurse (rn) reported that a dialyzer leaked immediately after the initiation of the patient¿s hemodialysis (hd) treatment.The rn visually observed that the dialyzer leaked from two locations.The first leak occurred two inches below the header cap of the dialyzer.The second leak occurred within a fiber.The dialyzer leaks were confirmed to be internal.The machine, a fresenius 2008t machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were not used to test for the presence of blood.There was no defect or damage seen on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 100 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on a new machine and treatment completed successfully with new supplies.The complaint device was reported to be available to be returned to the manufacturer for evaluation.
|
| |
|