Catalog Number UNKNOWN |
Device Problems
Structural Problem (2506); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Chest Pain (1776); Dyspnea (1816); Internal Organ Perforation (1987); Perforation of Vessels (2135); Insufficient Information (4580)
|
Event Type
Injury
|
Manufacturer Narrative
|
Reporter occupation: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
|
|
Event Description
|
It is alleged that the patient received a gunther tulip on (b)(6) 2007.It is alleged that the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
|
|
Manufacturer Narrative
|
Investigation: the following allegations have been investigated: aortic perforation, chest pain, shortness of breath.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Unknown if the reported chest pain and shortness of breath are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
The patient alleged aortic perforation.The patient further alleged severe and persistent chest pain, and shortness of breath.(b)(6) 2019, per a report from computed tomography; ¿impression: ivc filter anchor has penetrated the medial wall of the ivc and is embedded in the wall of the descending abdominal aorta.¿.
|
|
Manufacturer Narrative
|
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: organ perforation.The additional information regarding organ perforation does not change the previous investigation results for aortic perforation.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications, and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56 this report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
Additional information received alleges organ perforation.
|
|