• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC ADVANCED ENERGY (SALIENT) PLASMABLADE 3.0S; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC ADVANCED ENERGY (SALIENT) PLASMABLADE 3.0S; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number PS210-030S
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare professional (hcp) via manufacture representative (rep) regarding a generator and a handpiece.It was reported that the surgeon was using the handpiece during a surgical procedure and the black protective shield at the tip of the hand piece fell-off.A new handpiece was used to complete the procedure.No other information was provided.There was no impact to patient outcome.
 
Additional Manufacturer Narrative
H3: analysis summary: complaint confirmed: the blade was observed as well and showed the coating were considerably worn when it was used.With the observations, it was confirmed the device did not fail from manufacturing but from inadequate maintenance to the device as recommended in the ifu.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLASMABLADE 3.0S
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
MEDTRONIC ADVANCED ENERGY (SALIENT)
180 international drive
portsmouth NH 03801
MDR Report Key11543629
Report Number1226420-2021-00066
Device Sequence Number4602508
Product Code GEI
UDI-Device Identifier00643169617148
UDI-Public00643169617148
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K093695
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/28/2022
Device Model NumberPS210-030S
Device Catalogue NumberPS210-030S
Device Lot Number0218773537
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2021
Supplement Date Received by Manufacturer06/28/2021
Initial Report FDA Received Date03/22/2021
Supplement Report FDA Received Date06/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-