Model Number 288022
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Investigation summary: the device was received and evaluated at the service center.The reported complaint that the device doesn't turn due to a jammed motor, was confirmed.It was found that the motor was sticking as it was corroded.Further, the cable resistance of the motor cable was out of range.The motor and the motor cable were replaced and the device was repaired, tested and found to be fully functional.Fluid ingress into the system and contact with the motor is responsible for the corrosion of the motor.The corroded motor has a tendency to stick and not turn, hence is responsible for the issue reported by the customer.However, given the information provided, we cannot determine a definitive root cause for the defective motor cable.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 09/16/2019 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the customer in (b)(6) that during an unknown surgery on an unknown date, it was observed that the handpiece device did not turn as it had a jammed motor.During in-house engineering evaluation, it was determined that the motor was sticking as it was corroded.Another like device was used to complete the procedure without delay.There were no adverse patient consequences reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8022 HANDPC TORNADO SHAVER
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key11544390
MDR Text Key245124592
Report Number1221934-2021-00913
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705018345
UDI-Public10886705018345
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number288022
Device Catalogue Number288022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-