U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PANTERA LEO 2.5/12; BASIC CORONARY ANGIOPLASTY BALLOON CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PANTERA LEO 2.5/12; BASIC CORONARY ANGIOPLASTY BALLOON CATHETER Back to Search Results
Model Number 367004
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
A pantera leo coronary balloon catheter was selected for treatment of a severely calcified lesion (93 percent stenosis degree) in the severely tortuous mid lad.Pressure was applied (20 atm) but the balloon did not inflate properly.Contrast medium leaked from balloon.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned instrument revealed a small tear in the balloon about 3.5 mm distal to the proximal x-ray marker.Scratches were observed in the balloon surface in close vicinity to the tear.The damage of the balloon surface was most likely caused by a hard, sharp-edged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy (i.E.Severe calcification).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANTERA LEO 2.5/12
Common Device Name
BASIC CORONARY ANGIOPLASTY BALLOON CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key11544570
Report Number1028232-2021-01362
Device Sequence Number3147781
Product Code LOX
UDI-Device Identifier07640130415621
UDI-Public07640130415621
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K163660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model Number367004
Device Catalogue NumberSEE MODEL NO.
Device Lot Number08203082
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2021
Initial Date Received by Manufacturer 03/22/2021
Supplement Date Received by Manufacturer05/18/2021
Initial Report FDA Received Date03/22/2021
Supplement Report FDA Received Date05/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-