| Model Number |
367004 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A pantera leo coronary balloon catheter was selected for treatment of a severely calcified lesion (93 percent stenosis degree) in the severely tortuous mid lad.Pressure was applied (20 atm) but the balloon did not inflate properly.Contrast medium leaked from balloon.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned instrument revealed a small tear in the balloon about 3.5 mm distal to the proximal x-ray marker.Scratches were observed in the balloon surface in close vicinity to the tear.The damage of the balloon surface was most likely caused by a hard, sharp-edged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy (i.E.Severe calcification).
|
| |
|