ARTHREX, INC. FIBULOCK IMPLANT SYSTEM, STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number FIBULOCK IMPLANT SYSTEM, STERILE
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that the long flexible wire from the ar-8973ds fibulock implant system was introduced and advance up the fibula.When the 3.2mm reamer was introduced and advanced up the fibula, the device came to a hard stop.When attempting to advance the 3.2mm reamer up the fibula further, the flexible wire that was being reamed over broke in half.The broken flexible wire was left in the fibula and the ar-8973l-30-130 fibulock nail was inserted per technique guide.The proximal tip of the fibulock nail came into contact with the broken end of the wire and as the fibulock nail passed by the broken wire, the proximal tip of the fibulock bent.The fibulock nail and flexible wire were both left in the fibula as the fracture was reduced and the doctor did not have a problem leaving it in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBULOCK IMPLANT SYSTEM, STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key11547674
Report Number1220246-2021-02771
Device Sequence Number16994855
Product Code HSB
UDI-Device Identifier00888867303041
UDI-Public00888867303041
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K173656
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 03/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFIBULOCK IMPLANT SYSTEM, STERILE
Device Catalogue NumberAR-8973DS
Device Lot Number10763386
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2021
Initial Report FDA Received Date03/22/2021
Patient Sequence Number1
-
-