| Model Number |
7209497 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, while suturing the soft tissues in the shoulder, the surgeon saw that the branch of the arthropierce device did not lock.When it was removed from the joint, it was noticed that the jammed branch broke off.A backup device was available to complete the procedure with no significant delay or other complications.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.An analysis of the customer provided images found that the upper jaw of the device had broken and separated from the assembly.The batch number was confirmed.A visual inspection confirmed that the upper jaw had broken off.The broken pieces were returned in a small envelope.There was some wear on the device, but no debris.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.The complaint was confirmed and the root cause was associated with component failure.Factors that could have contributed to the reported event include more frequent use than anticipated for the device, rough sterilization processes, or inability to withstand excessive forces during use.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|