Catalog Number UNKNOWN
Device Problems Structural Problem (2506); Insufficient Information (3190)
Patient Problems Pain (1994); Perforation of Vessels (2135); Discomfort (2330); Insufficient Information (4580)
Event Type  Injury  
Manufacturer Narrative
It has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog # is unknown but referred to as gunther tulip.Occupation: non-healthcare professional.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Original.It is alleged that the patient received a gunther tulip on (b)(6) 2017.It is alleged that the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
 
Event Description
Patient allegedly received an implant on (b)(6) 2017 via the right common femoral vein due to pulmonary embolism.Patient is alleging vena cava perforation.Patient further alleges pain and discomfort.Report from computerized tomography (ct): "an ivc filter is present.The ivc filter has been placed at the renal arteries with approximately 15mm of the upper portion being above the renal arteries.The larger tines protrude outside of the ivc but do not contact any vital structures.No broken tine.No displaced tine.".
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: a2, b1, b5, b6, b7, h6.Additional information: investigation.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: vena cava perforation, pain, discomfort.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Unknown if the reported pain and discomfort are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown; however, there is no evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
MDR Report Key11547877
MDR Text Key245849651
Report Number3002808486-2021-00996
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/01/2021
Initial Date FDA Received03/22/2021
Supplement Dates Manufacturer Received07/14/2021
Supplement Dates FDA Received08/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight118
-
-