SMITH & NEPHEW, INC. GII DISHED INS SZ 3-4 9MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CNSTRND, CEMENTED, POLY/METAL/POLY
|
Back to Search Results |
|
| Model Number |
71420766 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during total knee arthroplasty the genesis ii dished ins size 1-2 9mm and the genesis ii dished articular insert size 3-4 mm would not seat into tibia tray.No delay or injury reported.Backup devices from smith and nephew were opened and seated correctly.
|
| |
|
Additional Manufacturer Narrative
|
|
The device, used in treatment, was returned for evaluation.A visual inspection of the returned insert confirms minor damage to the lock detail, more than likely from attempting to snap into the tibia tray.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Some potential probable causes for this event could include a fit/ sizing issue or poor insertion technique.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|