| Model Number |
SRLP-400 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade iii.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: weight to the spec, red particles and opening.A microscopic analysis was performed which identify striated opening.Based on the device analysis the final assessment is: a striated edge opening on anterior side assessed as surgical damage.
|
| |
|
Event or Problem Description
|
|
Healthcare professional additionally reported right side rupture.
|
| |
|
Search Alerts/Recalls
|