OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12, STERILE, SINGLE USE, 12 PC; HF-RESECTION ELECTRODES Back to Search Results
Model Number A22201C
Medical Device Problem Codes Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event or Problem Description
Olympus was informed that during a therapeutic transurethral resection of a bladder tumor (turbt) procedure, the loop wire at the distal end of the hf resection electrode broke and fragments fell into the patient.It is unknown whether or not fragments may have remained inside the patient.An x-ray examination was to be performed to confirm that no fragments remained inside the patient, the results of which have not been provided.However, the intended procedure was completed using a similar device and there was no report about an adverse event or patient injury.
 
Additional Manufacturer Narrative
The suspect medical device was returned to the manufacturer for investigation.The investigation confirmed that the loop wire at the distal end of the resection electrode is broken and the loop burnt down subsequently.During the announced x-ray examination no parts were found inside the patient.The cause of this event was attributed to wear and tear during use.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12, STERILE, SINGLE USE, 12 PC
Common Device Name
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
MDR Report Key11552490
Report Number9610773-2021-00103
Device Sequence Number11719541
Product Code FAS
UDI-Device Identifier14042761036648
UDI-Public14042761036648
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K931763
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 06/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/02/2025
Device Model NumberA22201C
Device Catalogue NumberA22201C
Device Lot Number1000059075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/18/2021
Supplement Date Received by Manufacturer05/27/2021
Initial Report FDA Received Date03/23/2021
Supplement Report FDA Received Date06/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No